资源描述
Click to edit Master title style,Click to edit Master text styles,Second level,Third level,Fourth level,Fifth level,M4735A,操作培训,*,Click to edit Master title style,Click to edit Master text styles,Second level,Third level,Fourth level,Fifth level,Philips healthcareM4735A,1,、,M4735A,2,、主机和配件的构成与连接,3,、主要功能和操作,M4735A,构造介绍,-,后视图,打印机,ECG,输入,电源,ECG,输出,SPO2,M4735A,配件连接,M4735A,主界面,除颤,自动检测,/,充电,同步,/,非,起搏模式,功能键,4,功能键,6,心电波幅,功能键,5,起搏,QRS,波音量,事件概要,记录,注释,调整起搏电流,调整起搏率,功能键,7,扬声器,M4735A,除颤手柄构造介绍,接触指示器,儿童极板,儿童极板,-,锁扣,M4735A,导联选择,除颤第,1,步,能量选择,除颤第,2,步,充电,手柄黄色按钮,面板,PACER,1.,如设备电源已接通,转动“能量选择”旋钮到“,Off,”(断)位置。,2.,同时按下软键,4,与,5,,转动“能量选择”旋钮到“,AED On,”(,AED,通)位置。出现配置菜单,如图 所示。菜单列出可配置的设置种类。,General Settings,AED Settings,Manual Settings,ECG Filter Settings,Save Settings to Data Card,Load Settings from Data Card,Print All Settings,Turn Off To Exit Configuration,MAIN,ENTER,每日检测,+AED ON,可以适用于医院对于急救设备的规定,也可检查设备的运行情况,除颤仪维护及保养,显示屏出现,low battery(,电池低电压,),提示后,除颤器可再提供,30,分钟的监护和,5,次,200J,的除颤,1,个月以上不用除颤时,应将电池取出,并在机器上注明电池已被取走,清洁,请使用肥皂水清洁监护仪附件,尽量不要使用酒精或其它腐蚀性强的清洁剂。,不推荐在此监护仪、相关的产品、附件式供应品上进行消毒。,除颤仪打印头可用软布蘸酒精清洗,
展开阅读全文