资源描述
水质分析仪表设计参数目录
在线水质分析仪
PH在线分析仪````````````````````````````````````````````````````````````````````````````````````````````6
溶解氧分析仪``````````````````````````````````````````````````````````````````````````````````````````````7
ORP在线分析仪``````````````````````````````````````````````````````````````````````````````````````````9
电导率在线分析仪````````````````````````````````````````````````````````````````````````````````````````10
污泥浓度/浊度分析仪````````````````````````````````````````````````````````````````````````````````````11
COD在线分析仪``````````````````````````````````````````````````````````````````````````````````````````13
总磷在线分析仪```````````````````````````````````````````````````````````````````````````````````````````14
氨氮在线分析仪```````````````````````````````````````````````````````````````````````````````````````````15
污泥界面仪`````````````````````````````````````````````````````````````````````````````````````````````````16
污泥体积指数分析仪`````````````````````````````````````````````````````````````````````````````````````17
明渠流量计`````````````````````````````````````````````````````````````````````````````````````````````````18
自动水质采样器```````````````````````````````````````````````````````````````````````````````````````````19
浊度分析仪`````````````````````````````````````````````````````````````````````````````````````````````````20
余/总氯分析仪`````````````````````````````````````````````````````````````````````````````````````````````21
二氧化氯分析仪```````````````````````````````````````````````````````````````````````````````````````````22
硬度分析仪`````````````````````````````````````````````````````````````````````````````````````````````````23
碱度分析仪`````````````````````````````````````````````````````````````````````````````````````````````````23
实验室水质分析仪
便携式溶解氧分析仪(HQ10)````````````````````````````````````````````````````````````````````````24
便携式pH/ORP分析仪(senSION1)````````````````````````````````````````````````````````````````24
便携式电导分析仪(senSION5)``````````````````````````````````````````````````````````````````````24
便携式污泥浓度/污泥界面仪(HS1)````````````````````````````````````````````````````````````````25
实验室COD/总磷/总氮分析仪(DR2500分光光度计+DRB数字式消解器)```````````````26
实验室BOD分析系统(生物培养箱+BODTrak)````````````````````````````````````````````````29
便携式磷酸盐分析仪(袖珍比色计II)``````````````````````````````````````````````````````````````30
便携式氨氮分析仪(袖珍比色计II)`````````````````````````````````````````````````````````````````30
实验室浊度计```````````````````````````````````````````````````````````````````````````````````````````````31
便携式余氯/二氧化氯分析仪````````````````````````````````````````````````````````````````````````````32
二、 仪表系统
§1 总述
原则上全部采用进口技术先进的数显式电动仪表,必须是国际公认成熟可靠的品牌,输出信号为标准4~20mA DC信号,消耗品根据需要提供相应的备品备件。
§2 自控仪表技术要求
2.1 基本要求
集成商所选用的仪表必须是成套配备,包括仪表本身及所有安装所需的各种附件以及连接线。仪表系统是用来连续测量污水处理工艺流程中的主要参数,并将测量数据送入计算机数据采集及监控系统。
室外安装设备必须加装防护罩/箱。所有仪表自控设备必须考虑电源及信号配置防雷、防感应电流冲击设施。设备的连接电缆均应穿管或沿电缆桥架敷设,出桥架端必须穿管,不得有露头在外。
所有设备的安装调试应
合相关的GB/GT要求,无相关GB/GT时符合厂家安装调试要求。
仪表的安装管件必须与连接管道相协调,统一标准。
仪表除内部专用电线外其它或与控制器间的信号等应采用屏蔽电缆。
仪表自控安装时需要的安装附件、连接件及开关电源等应包括在报价内。
2.2 环境
正常环境温度:-10℃~+50℃
相对湿度:0~95% (无冷凝)
海拔高度:<500m
所有检测元件,变送器,安装支架及保护罩等材料应满足污水环境防腐蚀要求。
在该环境下,所有设备必须能够正常工作,并达到文件所规定的可靠性和精度。
2.3工作电源
所有设备应由以下一种电源供给:
220VAC±10%,50HZ±1HZ;
2.4 防护等级
机箱设备外壳等级严格按照IEC529标准执行。
室内地面上设备等级≧IP54。
安装在水下或其它类似地区的设备等级IP68。
2.5 信号接口
所有仪表应采用以下信号电平:
仪表的信号:采用4~20mA直流电流信号
2.6 其他
(1)应提供所需的辅助和附加装置。例:检测元件至变送器电,变送器安装支架,保护罩、清洗所用空压机等。
(2)应提供厂家可靠的设计产品。
(3)产品应易于维护和检查,并提供易损害的备用部件。
(4)应符合国际通用标准和规范。
(5)应有零度和满度调整电路,并能方便地进行调整。
(6)应设计有温度补偿电路和抗干扰电路。
(7)检测精确度和响应时间应满足污水厂工艺和自控要求。
(8)应提供仪表维修保养用专用工具二套。
(9)应提供仪表专用的试剂及电解液(三年使用)
(10)★系统集成商所供设备不得低于选型参考所列厂家设备的档次。所供设备的测量范围必须与标书要求一致或涵盖标书要求的范围。
(11)仪表信号通过屏蔽电缆和PLC采集模块连接
§3 仪表安装地点,规格及数量
在此规定的供货范围包括下列仪表及其标书中所述的附属部件,以及总则中规定的各项服务
§4 供货范围
★总则:标准化的外观、运行、维修、备品备件以及制造商服务,所提供的设备必须是一个制造商的最终产品。
应提供所需的辅助和附加装置。例:检测元件至变送器电缆,变送器安装支架,保护罩等,应提供厂家可靠的设计产品,产品应易于维护和检查,并提供易损害的备用部件,应提供仪表维修保养用专用工具二套。
★每台设备仪表应成套地配备安全、有效及可靠运行所需的附件。供货范围包括该设备所有部件的供应及概述及总则规定的各项服务不论本标书是否指明,设备必须的部件也由系统集成商负责,由系统集成商根据不同规格的设备数量决定备品规格,并在投标中给出所需三年的备品备件价格单列后计入总投标价中。
§5 材质
除技术文件中明文规定外,系统集成商应根据设备的使用环境,合理确定设备各部门的材质,以保证系统正常使用,并在投标书中标明各部件材质。
本包所有招标设备的全部材料应适用于污水腐蚀环境。设备中的各部件应进行合理的防腐处理。具体作法由供货商确定,投标书中应予说明。
§6 资料提交
(1)投标时应提供的资料:
供方应提供下列资料(中文,8套)但不限于以下内容:
·设备的详图及说明,表明所有的安装尺寸及材料;
·详细说明设备的性能;
(2)中标后应提供的资料:
供方应提供下列资料(中文8套,电子文档2套)但不限于以下内容:
·安装方案的详细描述及安装精度规定;
·设备各部分的重量、平台土建荷载。
·安装、维修和运行手册;
·制造及质量保证设施;
·试运转前的调试及检测要求;
·上述设备装配图配套箱柜的接线图及原理图。
§7 制造商的服务
仪表制造商的技术代表到工作现场进行工作,协助安装、检查,提交安装完毕的证明。协助进行调试运行,进行启动前的培训和工作现场培训。
设备检验、验收及业主员工的培训将按照费用计入投标总价中,并应在投标书中单独列出。
§8 经验及资格
制造商应有10年以上的该类设备制造经历,并出示业绩表和等同规格设备的用户证明2个以上。
哈希水质分析仪表技术参数
北京安恒测试技术有限公司 0451--55500243 55500224 姜鹏 13936683220
(一)在线水质分析仪器
pH (P53+6028P0) 经济性配置
用途:测量、显示和传输污水处理过程中的pH值和温度。
探头:
(1)★电极:差分电极,带双阶参比电极(接地电极和参比电极),自动温度补偿;
(2)★测量范围:0~14pH/-5~95℃;
(3)★灵敏度:≤0.005pH;
(4)稳定性:每 24 小时 0.03 pH,不累积;
(5)★探头最大传输距离:914米;
(6)防护等级:IP68;
(7)电缆线长:3米;
(8)安装方式:浸没式;
变送器:
(1)工作环境:-20-60℃,0~95%相对湿度,无冷凝;
(2)★精度:0.1%;
(3)重现性:0.1% ;
(4)★温度补偿范围:-10~110℃,自动温度补偿;
(5)★显示:LCD液晶显示屏,可现场显示PH测量值和温度值,中文操作菜单;
(6)输出:两路0/4~20mA输出(pH和温度),继电器;
(7)通讯:RS-232;
(8)防护等级:NEMA4X,IP66;
(9)电源:180~260VAC,50/60Hz;
(10)安装方式:壁挂/面板/管道式安装;
附件:
(1) 探头浸没式安装支架;
(2) 空气自清洗系统。
pH (PHD+SC100) 主力型配置
用途:测量、显示和传输污水处理过程中的pH值和温度。
探头:
(1)★电极:差分电极,带双阶参比电极(接地电极和参比电极),自动温度补偿;
(2)★测量范围:0~14pH/-5~95℃;
(3)★灵敏度:≤0.005pH;
(4)稳定性:每 24 小时 0.03 pH,不累积;
(5)★探头最大传输距离:标准100米,最远达1000米;
(6)防护等级:IP68;
(7)电缆线长:4.5米;
(8)安装方式:浸没式;
变送器:
(1) 工作环境:-20-60℃,0~95%相对湿度、无冷凝;
(2) ★准确性:±0.1%;
(3) 灵敏度:±0.05%;
(4) 重现性:±0.05%;
(5) ★温度漂移:零点和范围:整个量程范围内为±0.02%/℃;
(6) 响应时间:90%的测量值的响应时间为 1-60秒;
(7) ★显示:LCD液晶显示屏,具数据存贮功能,具有中文菜单显示;
(8) ★探头输入:双通道,探头可即插即用,无需校准;
(9) ★输出信号:两路0/4~20mA DC,三个继电器输出;
(10) 通讯协议:MODBUS、Profibus DP (选配);
(11) 存储器备份:用户设置均保存在存储器中;
(12) 防护等级:NEMA4X;
(13) 电源:180 ~ 260VAC,50/60Hz;
(14) 安装方式:壁挂/面板/管道式安装;
溶解氧分析仪(LDO+SC100) 主力型配置
用途:测量、显示和传送污水生化处理过程溶解氧浓度。
探头:
(1) ★形式:无膜、无阴阳电极、无电极液,抗H2S、金属离子、油污染;
(2) ★工作原理:化学荧光法;
(3) ★溶氧测量范围:0~20 mg/L,0~20ppm或0-200%饱和度;
(4) ★温度测量范围:0~50℃,带温度自动补偿;
(5) 分辨率:0.01 mg/L;
(6) 重现性:;0.05mg/L
(7) 灵敏度:0.05mg/L;
(8) ★响应时间:20°C时,达到90%<30秒,达到95%<90秒;
(9) 防护等级:IP68;
(10) 电缆:10米,带快速接头;
变送器:
(15) 工作环境:-20-60℃,0~95%相对湿度、无冷凝;
(16) ★量程:溶氧:0~20mg/L,温度:0~100℃;
(17) ★准确性:±0.1%;
(18) 灵敏度:±0.05%;
(19) 重现性:±0.05%;
(20) ★温度漂移:零点和范围:整个量程范围内为±0.02%/℃;
(21) 响应时间:90%的测量值的响应时间为 1-60秒;
(22) ★显示:LCD液晶显示屏,具数据存贮功能,具有中文菜单显示;
(23) ★探头输入:双通道,探头可即插即用,无需校准;
(24) ★输出信号:两路0/4~20mA DC,三个继电器输出;
(25) 通讯协议:MODBUS、Profibus DP (选配);
(26) 存储器备份:用户设置均保存在存储器中;
(27) 防护等级:NEMA4X;
(28) 电源:180 ~ 260VAC,50/60Hz;
(29) 安装方式:壁挂/面板/管道式安装;
附件:
(1) 探头浮球式安装组件;
(2) 空气自清洗系统。
Application :Aeration Tanks, Collection systems, Nitrification & denitrification tanks, Anaerobic Digesters, Aerobic Digesters, Up and Downstream from plant effluent (NPDES Permit Monitoring),Plant outfall(NPDES Permit Monitoring)
Sensor: 1. No Membranes, No Electrolyte solutions to poison or replenish, No anode or cathode to clean or replace.
2.No Electrodes, Sensor is coated with a luminescent material.
3. Measuring Range: Unlimited between 0—20.00mg/L, 0.00-20.00ppm, 0-200% Saturation
4. Temperature Range: 0-50ºC
5. Resolution: 0.01ppm 0.01mg/L
6. Repeatability:0.05ppm
7. Sensitivity: +0.05% of Span
8. Response Time: To 90%<60 Seconds; To 95%<90Seconds
9. Protection degree :IP68
Transmitter: 1.Operation temperature :-20—60ºC
2.Measuring Range: 0.00-20.00ppm, 0.00—20.00mg/L, or0-99.9% Saturation
3.Accuracy: +0.1% of Span
4.Sensitivity: +0.05% of Span
5. Repeatability: +0.05ppm of Span
6. Temperature Drift: Zero and Span +0..02% of Span per ºC
7.Display:Graphic dot matrix LCD, 128x64 pixels with LED backlighting, with Chinese/English languages
8.Sensor to Transmitter Distance:300m(maximum)
9. Temperature Compensation: Automatic from 0.0—50.0ºC
10. Input: Duel sensor input
11. Output: Dual Isolated 0/4-20mA ,with RS232, HART Protocol, RS-485 MODBUS
12. Relays Types/Output: Three electromechanical relays SPDT
ORP 氧化还原电位检测仪(P53+2028R0) 经济型配置
用途:用于测量、显示和变送水解酸化阶段以及反硝化过程中的氧化还原电位,安装于生化池。
探头:
(1) ★差分式电极(含温度电极),带双阶参比电极(接地电极和参比电极),带温度补偿;
(2) ★测量范围:-2000~+2000mv/-5~95℃;
(3) ★灵敏度:≤0.5mV;
(4) 稳定性:每24小时2mV,不累积;
(5) 探头最大传输距离:914米;
(6) ★防护等级:IP68;
(7) 电缆线长:3米;
(8) 安装方式:浸没式;
变送器:
(1)工作环境:-20~60℃,0~95%相对湿度,无冷凝;
(2)★精度:0.1%;
(3)重现性:0.1% ;
(4)★温度补偿范围: -10~110℃,自动温度补偿;
(5)★显示:LCD液晶显示屏,可现场显示ORP测量值和温度值,中文操作菜单;
(6)输出:两路0/4~20mA输出(ORP和温度),继电器;
(7)★通讯:RS-232;
(8)防护等级:NEMA4X;
(9)电源:180~260VAC,50/60Hz;
(10)安装方式:壁挂/面板/管道式安装;
安装附件:
(1) 探头浸没式安装附件;
(2) 空气自清洗系统。
ORP 氧化还原电位检测仪(SC100+2028R0) 主力型配置
用途:用于测量、显示和变送水解酸化阶段以及反硝化过程中的氧化还原电位,安装于生化池。
探头:
(1) ★差分式电极(含温度电极),带双阶参比电极(接地电极和参比电极),带温度补偿;
(2) ★测量范围:-2000~+2000mv/-5~95℃;
(3) ★灵敏度:≤0.5mV;
(4) 稳定性:每24小时2mV,不累积;
(5) 探头最大传输距离:914米;
(6) ★防护等级:IP68;
(7)电缆线长:3米;
(8)安装方式:浸没式;
变送器:
(1) 工作环境:-20-60℃,0~95%相对湿度、无冷凝;
(2) ★准确性:±0.1%;
(3) 灵敏度:±0.05%;
(4) 重现性:±0.05%;
(5) ★温度漂移:零点和范围:整个量程范围内为±0.02%/℃;
(6) 响应时间:90%的测量值的响应时间为 1-60秒;
(7) ★显示:LCD液晶显示屏,具数据存贮功能,具有中文菜单显示;
(8) ★探头输入:双通道,探头可即插即用,无需校准;
(9) ★输出信号:两路0/4~20mA DC,三个继电器输出;
(10)通讯协议:MODBUS、Profibus DP (选配);
(11)存储器备份:用户设置均保存在存储器中;
(12)防护等级:NEMA4X;
(13)电源:180 ~ 260VAC,50/60Hz;
(14)安装方式:壁挂/面板/管道式安装;
电导率在线测量仪(E53+3700) 经济型配置
用途:测量、显示和变送自来水厂、工业过程用水和污水处理厂水处理过程的电导率。
探头:
(1) 测量原理:感应电流;
(2) ★传感器:无极非接触式,带PT1000温度传感器;
(3) 温度范围:-10~200℃;
(4) ★测量范围:0~200uS/cm到0~2,000,000mS/cm;
(5) 电缆线长:6米;
(6) 防护等级:IP68;
(7) 安装方式:浸没式或管道式;
变送器:
(1) 工作环境:-20-60℃,0~95%相对湿度、无冷凝;
(2) 准确性:±0.1%;
(3) 灵敏度:±0.05%;
(4) 重现性:±0.05%;
(5) 温度漂移:零点和范围:整个量程范围内为±0.02%/℃;
(6) 响应时间:90%的测量值的响应时间为 1-60秒;
(7) 显示:LCD液晶显示屏,具数据存贮功能,具有中文菜单显示;
(8) 探头输入:双通道,探头可即插即用,无需校准;
(9) 输出信号:两路0/4~20mA DC,三个继电器输出;
(10) 通讯协议:MODBUS、Profibus DP (选配);
(11) 存储器备份:用户设置均保存在存储器中;
(12) 防护等级:NEMA4X;
(13) 电源:180 ~ 260VAC,50/60Hz;
(14) 安装方式:壁挂/面板/管道式安装;
污泥浓度/浊度分析仪(Solitax sc) 主力型配置
用途:用于污水处理厂进出水、曝气池和二沉池中固体悬浮物和浊度的测量、显示和传送。
探头:
(1)★原理:双光束近红外光/散射光,90°和140°检测器,不受样品颜色干扰;
(2)★测量范围:固体悬浮物:0~150g/L,浊度:0~4000NTU;
(3)★精度:测量值的±1%;
(4)重现性:1%;
(5)★响应时间:<1秒,可调整;
(6)★形式:316不锈钢或PVC材质,具有自诊断功能和机械式刮片自清洗功能;
(8)工作温度:2℃~40℃;
(7)防护等级:IP68;
(8)电缆长度:8米;
(9)安装方式:浸没式或管道式;
变送器:
(1) 工作环境:-20-60℃,0~95%相对湿度、无冷凝;
(2) ★准确性:±0.1%;
(3) 灵敏度:±0.05%;
(4) 重现性:±0.05%;
(5) ★温度漂移:零点和范围:整个量程范围内为±0.02%/℃;
(6) 响应时间:90%的测量值的响应时间为 1-60秒;
(7) ★显示:LCD液晶显示屏,具数据存贮功能,具有中文菜单显示;
(8) ★探头输入:双通道,探头可即插即用,无需校准;
(9) ★输出信号:两路0/4~20mA DC,三个继电器输出;
(10) 通讯协议:MODBUS、Profibus DP (选配);
(11) 存储器备份:用户设置均保存在存储器中;
(12) 防护等级:NEMA4X;
(13) 电源:180 ~ 260VAC,50/60Hz;
(14) 安装方式:壁挂/面板/管道式安装;
污泥浓度检测仪(Txpro2+RD260), 经济性配置
用途:测量、显示和变送污水处理过程中的污泥浓度。
探头:
(1)★原理:4-beam交互四光束原理,自动补偿样品颜色变化;
(2)★测量范围:0-25g/L;
(3)★精度:±2%读数;
(4)重现性:±1%读数;
(5)★自清洗:空气自清洗系统;
(6)材质:CPVC;
(7)防护等级:IP68,具有自诊断功能;
(8)电缆线长:10米;
(9)安装方式:浸没式或管道式;
控制器:
(1)工作环境:-20-55℃;
(2)读数频率:0.5秒一次;
(3)★信号平均:1秒至5分钟,用户可调;
(4)★显示:LCD背光液晶显示;
(5)★输入:双探头;
(6)输出信号:两路0/4~20mA,继电器;
(7)防护等级:NEMA4X(IP66);
(8)电源:90-265 VAC,50/60 Hz;
(9)安装方式:面板或壁挂安装;
安装附件:
(1) 探头浸没式或管道式安装组件;
(2) 探头空气自清洗装置。
COD分析仪( UVSAC)
用途:测量、显示和变送污水处理厂进水和出水的COD浓度。
探头:
(1)★原理:双波长(254和550nm)紫外光吸收法,自动补偿浊度的影响,无需化学试剂;
(2)★测量范围:1~800mg/L CODUV;
(3)准确度:±10%;
(4)重现性:测量值的±2%;
(5)★测量周期:1~30分钟,用户可调;
(6)★自清洗:机械式刮片自清洗装置;
(7)材质:316不锈钢;
(8)工作温度:2~40℃;
(9)防护等级:IP68;
(10)电缆线长:8米;
(11)安装方式:浸没式或流通式;
变送器:
(1)工作环境:-20-60℃,0~95%相对湿度、无冷凝;
(3) ★准确性:±0.1%;
(4) 灵敏度:±0.05%;
(5) 重现性:±0.05%;
(6) ★温度漂移:零点和范围:整个量程范围内为±0.02%/℃;
(7) 响应时间:90%的测量值的响应时间为 1-60秒;
(8) ★显示:LCD液晶显示屏,具数据存贮功能,具有中文菜单显示;
(9) ★探头输入:双通道,探头可即插即用,无需校准;
(10) ★输出信号:两路0/4~20mA DC,三个继电器输出;
(11) 通讯协议:MODBUS、Profibus DP (选配);
(12) 存储器备份:用户设置均保存在存储器中;
(13) 防护等级:NEMA4X;
(14) 电源:180 ~ 260VAC,50/60Hz;
(15) 安装方式:壁挂/面板/管道式安装;
CODcr分析仪
(1) 测试方法:重铬酸钾高温消解,比色测定
(2) 测试量程:0~5,000mg/l
(3) 检测下限:3.3mg/l
(4) 分辨率:<1mg/l
(5) 准确度:>100mg/l:< 10% 读数;<100mg/l:<± 6mg/l
(6) 重现性:>100mg/l:< 5% 读数;<100mg/l:<± 5mg/l
(7) 响应时间(>90%) :20min
(8) 测试间隔:20,30,40,60,80,100,120min,自动
(9) 校正间隔:按选定间隔自动进行(持续时间:60min)
(10) 清洗间隔:按选定间隔自动进行(持续时间:10min)
(11) 服务间隔:24 个月
(12) 用户保养:保养间隔>1 个月,每月约1 小时
(13) 试剂消耗:约1 个月(试剂和标准液)
(14) 自我监测:自我监测泄漏;仪器状态自我诊断
(15) 模拟输出:2 路0/4~20mA 模拟输出
(16) 继电器控制:2 路24V 1A 继电器高低点控制
(17) 服务接口:RS232
(18) 可选配BUS:MODBUS RS485,Profibus DP
(19) 显示:大屏幕LCD 图表显示,240*128
(20) 操作菜单:中文/ 英文
(21) 数据存储:2,000 组
(22) 典型应用:工业污水
(23) 样品PH:1~12
(24) 样品压力:开放槽,无压
(25) 消解温度:175℃
(26) 工作温度:+ 5℃~+40℃
(27) 电源:230 V AC ± 10% / 50-60 Hz
(28) 功耗:约100 VA
(29) 尺寸:550 mm x 810 mm x 390 mm
(30) 重量:约22Kg(未接药剂)
总磷分析仪
用途:用于污水处理过程进出水总磷浓度的测量、显示和传送。
仪器:
(1)★原理:钼蓝法,具有自动校正、自动清洗、自动气、液传输技术,可以同时测量正磷酸盐的浓度;
(2)★测量范围:0.01~5.0 mg/L;
(3)★测量精度:±2%;
(4)测量周期:10分钟;
(5)★仪器具有自动校正、自动清洗、自动背景值修正功能;
(6)显示:LCD液晶屏,数字及图形显示,具数据存储功能;
(7)输出:两路0/4~20mA,继电器;
(8)★服务端口:RS232;
(9)★通讯协议:MODBUS,PROFIBUS(选配);
(10)工作温度:5~40℃;
(11)防护等级:IP54,室内安装;
(12)电源:230±10%VAC,50Hz;
附件:
(1)★三年正常使用所需要的消耗品及试剂;
(2)★水样预处理系统:超声波匀样化样品,自清洗形式,对水样进行采集处理,为分析仪提供适当的样品。
正磷酸盐分析仪(Phosphax Compact)
用途:用于污水处理过程好氧段和出水中正磷酸盐浓度的测量、显示和传送。
仪器:
(1)★原理:钒钼黄比色法;
(2)★测量范围:0.1~10.0 mg/L PO4-P;
(3)最低检出限:0.1mg/L;
(4)★测量精度:±1.5%或0.1mg/L;(手册与宣传资料不符)
(5)测量周期:10,15,20或30分钟,用户可调;
(6)★仪器具有自动校正、自动清洗、自动背景值修正功能;
(7)★显示:LCD液晶屏,数据及图形显示,具数据存储功能;
(8)输出:两路0/4~20mA,继电器;
(9)★服务端口:RS232;
(10)★通讯协议:MODBUS,PROFIBUS(选配);
(11)操作环境:10~40℃,90%相对湿度;
(12)防护等级:IP54,室内安装;
(13)电源:100~230VAC,50/60Hz;
附件
(1) ★附件:一年正常使用所需要的消耗品及试剂;
(2) ★水样预处理系统:膜过滤,自清洗形式,对水样进行采集处理,为分析仪提供适当的样品。
氨氮分析仪
用途:用于污水处理厂进出水氨氮浓度的测量、显示和传送。
仪器:
(1)★原理: 氨气逐出比色测量原理,分析仪具有自动气、液传输技术,具有自动校正、自动清洗、自动恒温功能,连续在线监测氨氮;
(2)★测量范围:0.2~12mg/L,2-120 mg/L,20-1200 mg/L NH4-N(量程由试剂确定);
(3)★准确度:测量值的±2.5%;
(4)最低检测限:0.2mg/L ;
(5)★测量周期:13、15、20或者30分钟(程序设定);
(6)样品流速要求:最小为0.1 l/h;
(7)★仪器具有自动校正、自动清洗;
(8)★显示:LCD液晶屏,数据和图形显示,具数据存储功能;
(9)输出:两路0/4~20mA,继电器;
(10)★服务端口:RS232;
(11)通讯协议:MODBUS,PROFIBUS(选配);
(12)防护等级:IP54,室内安装;
(13)电源:100~230VAC,50/60Hz;
附件
(1) ★一年正常使用所需要的消耗品及试剂;
(2) ★水样过滤系统:膜过滤,自清洗形式,对水样进行采集处理,为分析仪提供适当的样品。
硝氮分析仪( UV硝氮)
用途:用于污水处理厂生化处理过程中硝氮浓度的测量、显示和变送。
探头:
(1) ★原理:双波长(210nm和350nm)紫外光吸收技术测量水中亚硝酸盐和硝酸盐含量,测量光程为2mm,无需化学试剂;
(2) ★补偿:自动补偿浊度和有机物干扰;
(3) 工作温度:2~40℃;
(4) ★测量范围:0~50mg/L NO2-N + NO3-N;
(5) 准确度:±0.5mg/L (标准溶液)或读数值的±3%;
(6) 响应时间:>1分钟,可调整;
(7) ★自清洗:机械式刮片自清洗装置;
(8) 材质:316不锈钢;
(9) 防护等级:IP68;
(10) 电缆线长:8米;
(11) 安装方式:浸没式或流通式;
变送器:
(1)工作环境:-20-60℃,0~95%相对湿度、无冷凝;
(3) ★准确性:±0.1%;
(4) 灵敏度:±0.05%;
(5) 重现性:±0.05%;
(6) ★温度漂移:零点和范围:整个量程范围内为±0.02%/℃;
(7) 响应时间:90%的测量值的响应时间为 1-60秒;
(8) ★显示:LCD液晶显示屏,具数据存贮功能,具有中文菜单显示;
(9) ★探头输入:双通道,探头可即插即用,无需校准;
(10) ★输出信号:两路0/4~20mA DC,三个继电器输出;
(11) 通讯协议:MODBUS、Profibus DP (选配);
(12) 存储器备份:用户设置均保存在存储器中;
(13) 防护等级:NEMA4X;
(14) 电源:180 ~ 260VAC,50/60Hz;
(15) 安装方式:壁挂/面板/管道式