收藏 分销(赏)

EZ-TOC II分析仪实操说明.pdf

上传人:wei****ing 文档编号:4255312 上传时间:2024-08-30 格式:PDF 页数:64 大小:2.21MB
下载 相关 举报
EZ-TOC II分析仪实操说明.pdf_第1页
第1页 / 共64页
EZ-TOC II分析仪实操说明.pdf_第2页
第2页 / 共64页
EZ-TOC II分析仪实操说明.pdf_第3页
第3页 / 共64页
EZ-TOC II分析仪实操说明.pdf_第4页
第4页 / 共64页
EZ-TOC II分析仪实操说明.pdf_第5页
第5页 / 共64页
点击查看更多>>
资源描述

1、BA424C/07/en/06.0971085882Operating InstructionsEZ-TOC II analyzer CA52TOCAnalyzer for continuous real-time TOC measurement in water and wastewaterEndress+HauserEZ-TOC II analyzer CA52TOCEndress+Hauser3Table of contents1 Safety instructions.41.1 Designated use .41.2 Installation,commissioning and op

2、eration .41.3 Operational safety .41.4 Return.51.5 Notes on safety icons and symbols.52 Identification.62.1 Nameplate .62.1.1 Product structure .62.2 Scope of delivery.72.3 Certificates and approvals .73 Installation.83.1 Incoming acceptance,transport,storage.83.2 Installation conditions .83.2.1 Loc

3、ation .83.2.2 Compressed air supply.93.3 Installation instructions .103.3.1 Mounting .103.3.2 Installing the UV lamp.103.3.3 Connecting the gas sources .113.3.4 Connecting the medium .123.3.5 Preparing chemicals.133.4 Post-installation check.174 Wiring .184.1 Electrical connection.184.1.1 Preparing

4、wiring .184.1.2 Electrical connection of analyzer .194.1.3 Connecting the signal outputs .204.1.4 Connecting the relay outputs.214.2 Post-connection check.215 Operation.225.1 Operating elements.225.2 Display and keypad .236 Commissioning.246.1 Function check .246.2 Programming .246.2.1 COMMANDS menu

5、 .246.2.2 CALIBRATION menu .266.2.3 SETUP menu.296.2.4 DIAGNOSTICS menu .396.3 Communication .407 Maintenance.437.1 Biweekly maintenance.457.2 Monthly maintenance .457.2.1 Replenish reagents.457.2.2 Reset reagent timer .457.2.3 Inspect entire unit.457.3 Quarterly maintenance .477.3.1 Replacement of

6、pump tubing .477.3.2 Internal cleaning of analyzer .487.3.3 Leak-testing of gas lines.497.3.4 Manual liquid calibration.497.3.5 Automatic clean.507.3.6 Automatic calibration.517.3.7 Alarm Standards .518 Accessories.529 Troubleshooting .539.1 Messages.539.2 Spare parts.559.3 Return.589.4 Decommission

7、ing .589.5 Disposal.5810 Technical Data.5910.1 Input.5910.2 Output .5910.3 Power supply.6010.4 Performance characteristics .6010.5 Environment.6010.6 Process .6010.7 Mechanical construction .60Index.61Safety instructionsEZ-TOC II analyzer CA52TOC4Endress+Hauser1Safety instructions1.1Designated useCA

8、52TOC is an analyzer for continuous real-time TOC measurement in water and wastewater.The analyzer is particularly suitable for use in the following areas:Monitoring of organic carbon in water and wastewater Industrial processes Outlet monitoring in sewage treatment plants Monitoring of spring water

9、 and drinking water Control of methanol dosageAny other use than the one described here compromises the safety of persons and the entire measuring system and is not permitted.The manufacturer is not liable for damage caused by improper or non-designated use.1.2Installation,commissioning and operatio

10、nPlease note the following items:Installation,commissioning,operation and maintenance of the measuring system must only be carried out by trained technical personnel.Trained personnel must be authorized for the specified activities by the system operator.Electrical connection must only be carried ou

11、t by a certified electrician.Technical personnel must have read and understood these Operating Instructions and must adhere to them.Before commissioning the entire measuring point,check all the connections.Ensure that electrical cables and hose connections are not damaged.Do not operate damaged prod

12、ucts and secure them against unintentional commissioning.Mark the damaged product as being defective.Measuring point faults may only be rectified by authorized and specially trained personnel.If faults can not be rectified,the products must be taken out of service and secured against unintentional c

13、ommissioning.Repairs not described in these Operating Instructions may only be carried out at the manufacturers or by the service organization.1.3Operational safetyThe analyzer has been designed and tested to the highest standards and left the factory in perfect functioning order.Relevant regulation

14、s and European standards have been met.As the user,you are responsible for complying with the following safety conditions:Installation instructions Local prevailing standards and regulations.Immunity to interferenceThis instrument has been tested for electromagnetic compatibility in industrial use a

15、ccording to applicable European standards.Protection against interference as specified above is valid only for an instrument connected according to the instructions in these Operating Instructions.EZ-TOC II analyzer CA52TOCSafety instructionsEndress+Hauser51.4ReturnIf the analyzer has to be repaired

16、,please contact the responsible sales center.If the analyzer has to be returned,please return it cleaned to the responsible sales center.Please use the original packaging,if possible.Please enclose the completed Declaration of contamination(copy the second last page of these Operating Instructions)w

17、ith the packaging and the transportation documents.No repair without completed Declaration of contamination!1.5Notes on safety icons and symbolsElectrical symbols#Warning!This symbol alerts you to hazards that can cause serious damage to the instrument or to persons if ignored.Caution!This symbol al

18、erts you to possible faults which could arise from incorrect operation.They could cause damage to the instrument if ignored.!Note!This symbol indicates important items of information.%Direct Current(DC)A terminal at which DC is applied or through which DC flows.&Alternating Current(AC)A terminal at

19、which(sine-form)AC is applied or through which AC flows.)Ground connectingA terminal which,from the users point of view,is already grounded using a grounding system.*Protective ground terminalA terminal which must be grounded before other connections may be set up.bAlarm relayInputOutputDC voltage s

20、ourceTemperature sensorIdentificationEZ-TOC II analyzer CA52TOC6Endress+Hauser2Identification2.1NameplatePlease check the order code from the nameplate with the product structure and with your order.2.1.1Product structurea0010818Fig.1:Nameplate(example)Measuring rangeA0.015 to 10 mg/l TOCB0.1 to 100

21、 mg/l TOC(not with CA52TOC-*C*)C0.5 to 500 mg/l TOC(not with CA52TOC-*C*)D10 to 1000 mg/l TOC(not with CA52TOC-*C*)E50 to 5000 mg/l TOC(not with CA52TOC-*C*)F100 to 10000 mg/l TOC(not with CA52TOC-*C*)YSpecial version according to customer specificationSample transfer11 measuring point(not with CA52

22、TOC-*D*;CA52TOC-*E*)22 measuring pointsTIC removalAStandardBHighCUltra high(only with CA52TOC-A*)YSpecial version according to customer specificationPower supply0230 VAC,50/60 Hz1115 VAC,50/60 HzSample preparationANot selectedB1 x PA-2 PVC,1 to 8 m3/h(4.4 to 35 gpm)wastewaterC1 x PA-3 PVC,0.1 to 1.0

23、 m3/h(0.4 to 4.4 gpm)wastewaterD2 x PA-2 PVC,1 to 8 m3/h(4.4 to 35 gpm)wastewater(not with CA52TOC-*1*)E2 x PA-3 PVC,0.1 to 1.0 m3/h(0.4 to 4.4 gpm)wastewater(not with CA52TOC-*1*)YSpecial version according to customer specificationCarrier gas0To order separately1CO2 scrubber2CO2 scrubber+compressor

24、9Special version according to customer specificationOutputA0/4 to 20 mA+RS 232 unidirectionalCA52TOC-Complete order codeEZ-TOC IIAnaheim,CA 92807 USASerial No.:Mains:230 V,50/60 Hz300 VAIP54Meas.range:0.015 to 100 mg/l TOCOutput:0/4 to 20 mAAmbient temp.:0 to 40 C32 to 104 FContacts:4 relaysOrder Co

25、de:CA52TOC-A1B0A0ACA52TOC-1012/0C1000A517W0EZ-TOC II analyzer CA52TOCIdentificationEndress+Hauser72.2Scope of deliveryThe scope of delivery comprises:1 EZ-TOC II analyzer CA52TOC 1 set of tubing(4 pieces)1 set of fittings(17 pieces)2 bottles(5.5 US gal)3 bottles(2 US gal)5 bottle caps(with hole for

26、tube)Operating Instructions,English 1 Allen key 1 drain plug with gasket 1 special key for peristaltic pump 1 key for door lockIf you have any questions,please contact your supplier or your local sales center.2.3Certificates and approvalsDeclaration of conformityThe product meets the requirements of

27、 the harmonized European standards.It thus complies with the legal requirements of the EC directives.The manufacturer confirms successful testing of the product by affixing the 4 symbol.InstallationEZ-TOC II analyzer CA52TOC8Endress+Hauser3Installation3.1Incoming acceptance,transport,storage Make su

28、re the packaging is undamaged!Inform the supplier about any damage to the packaging.Keep the damaged packaging until the matter has been settled.Make sure the contents are undamaged!Inform the supplier about damage to the contents.Keep the damaged products until the matter has been settled.Check tha

29、t the order is complete and agrees with your shipping documents.The packaging material used to store or to transport the product must provide shock protection and humidity protection.The original packaging offers the best protection.Also,keep to the approved ambient conditions(see Technical data).If

30、 you have any questions,please contact your supplier or your local sales center.3.2Installation conditions3.2.1LocationThe analyzer can be installed in two ways:standalone on a table mounted on a wallThe analyzer requires a gravity drain which is vented to the atmosphere.If located inside a building

31、,a vent is required.The accumulation of halogens or other vapors is not permitted in this area.a0008101Fig.2:Dimensions663(26.10)610(24.00)572(22.50)558.9(22.00)529.7(20.85)508(20.00)365.4(14.39)369.2(14.54)?9.53(.375)305(12.00)841.9(33.14)802(31.57)762(30.00)mm(inch)EZ-TOC II analyzer CA52TOCInstal

32、lationEndress+Hauser93.2.2Compressed air supplyThe compressed air(used as a carrier gas)must be dry and must meet the following specifications:3 ppm CO2 5 l(1.3 US gal)a0008042Fig.4:Measuring system when air supply is stable1EZ-TOC II analyzer CA52TOC2Pressure regulator3CO2 scrubber4Stable air suppl

33、y(or bottle)6 bar(87 psi)123451243InstallationEZ-TOC II analyzer CA52TOC10Endress+Hauser3.3Installation instructions3.3.1MountingThe analyzer has four mounting holes in the back of the cabinet.Use at least 10 mm(5/16)mounting hardware to support the unit,as it is heavy.The mounting feet on the back

34、of the analyzer provide clearance between the wall and the cabinet.3.3.2Installing the UV lampCaution!The UV lamp is extremely fragile.Perform installation with care.a0011122Fig.5:UV reactorPreparation 1.Open the cabinet door and swing out the analyzers front panel.Block the door and panel,so that t

35、hey will not swing back.2.Remove the top Phillips screw on the reactor ground strap(pos.1,Fig.5).3.Remove the Phillips screw above the ground strap(pos.2,Fig.5).4.Remove the reactor cap.5.Remove the top of the reactor shell.12EZ-TOC II analyzer CA52TOCInstallationEndress+Hauser11Installing the lamp1

36、.Inside the reactor,remove the reactor nut and the two O-rings beneath it.2.Carefully slide first the reactor nut,then the two O-rings,over the glass end of the UV lamp to the top of the lamp.3.Insert the lamp into the center of the reactor.Exert a gentle downward pressure on the lamp as you tighten

37、 the reactor nut.The lamps metal sleeve must be against the nut.4.Caution!Tighten with care.Undertightening the reactor nut can result in leakage,causing damage to the analyzer.Overtightening the reactor nut can break the outer casing of the UV lamp.Hand-tighten the reactor nut,and then tighten it b

38、y one additional half turn.5.Replace the top,cap and ground strap on the reactor shell.3.3.3Connecting the gas sources Gas and liquid connections are all made via the connection panel located in the lower right-hand side of the cabinet.a0010722Fig.6:Connection panel for gas and liquid connectionsAll

39、 connections are made using PFA or Kynar fittings.Tighten the fittings securely,but be careful to avoid overtightening the fittings,as you may destroy the threads.CalibrationDeionizedWaterCleanDilutionSampleManualPersulfateSodiumAcidSample InSample OutDrainPurgeVentIR VentAir InSolution30 psiMax.Pre

40、ssure2 BarsInstallationEZ-TOC II analyzer CA52TOC12Endress+Hauser3.3.4Connecting the mediumConnecting the air supply:1.Use PFA tubing to attach the air supply line to the Air In fitting.2.Tighten the fittings securely.Caution!Do not under any circumstances allow the air pressure entering the analyze

41、r to exceed 2 bar(30 psi)maximum,as this could damage the analyzer and result in personal injury.3.Turn on the air supply.4.Use a mixture of water and soap to check for leaks.Paint the mixture directly onto the connections at the analyzer and the air supply.If there is a leak,bubbles will form aroun

42、d the connection.Tighten the fitting a little more and recheck the connection.Do not overtighten the fittings or you may destroy the threads.Connection of the liquid medium:1.Place the medium bottles close to the analyzer.2.Pour the reagents(acid,persulfate)into the two 5.5 US gal bottles.Pour the c

43、leaning,calibration and DI solutions into the three 2 US gal bottles.Use the bottle caps with a hole(included in the delivery)during operation of the analyzer.Use the bottle caps without a hole for transport and storage.3.Cut five pieces of 1/8 PFA tubing long enough to reach the bottom of the mediu

44、m bottles and the medium connections of the connection panel.4.Pass the ends of the tubes through the holes in the bottle caps to the bottom of the bottles.5.Connect the other ends of the tubes to the corresponding fittings on the connection panel.EZ-TOC II analyzer CA52TOCInstallationEndress+Hauser

45、133.3.5Preparing chemicals#Warning!Accidents with chemicals can kill,blind,burn,or damage lungs.Many chemicals are poisonous or corrosive;some are explosive,by themselves or in combination.Others are hazardous because they can be absorbed easily through skin or the respiratory system.When working wi

46、th any chemicals,always observe the precautions in the section Precautions regarding use of chemicals.Precautions regarding use of chemicals Read the material safety data sheets(MSDS)provided with each chemical to see what hazards are present and what precautions you must take.If you are not sure of

47、 the risks,seek assistance from qualified individuals.Never prepare chemicals by yourself.If an accident happens,you may need help.Make sure someone else is nearby.Prepare chemicals only in a properly-equipped lab facility.Always wear goggles,rubber gloves and a rubber apron when preparing chemicals

48、.If you work with finely powdered chemicals,wear a protective face mask or respirator.Never breathe,taste,or swallow any chemical or solution used with the EZ-TOC II analyzer(or any other process analytical device).When solutions are prepared,always label containers with contents and date of prepara

49、tion.Dispose of unlabeled or expired solutions in accordance with local regulations.Some chemicals react violently when combined or dissolved in water,causing explosions,fire,heat,or the dangerous discharge of gases,foam or spray.Avoid the risk of a serious accident;if you do not know how certain ch

50、emicals will react,do not combine them.Never combine chemicals known to react very strongly.There are three chemical solutions normally used with the EZ-TOC II.They are:Phosphoric acid,H3PO4,10%(v/v)solutionIt is used to acidify the sample solution.This transforms carbonate ion(CO3=)to CO2 and elimi

展开阅读全文
部分上传会员的收益排行 01、路***(¥15400+),02、曲****(¥15300+),
03、wei****016(¥13200+),04、大***流(¥12600+),
05、Fis****915(¥4200+),06、h****i(¥4100+),
07、Q**(¥3400+),08、自******点(¥2400+),
09、h*****x(¥1400+),10、c****e(¥1100+),
11、be*****ha(¥800+),12、13********8(¥800+)。
相似文档                                   自信AI助手自信AI助手
百度文库年卡

猜你喜欢                                   自信AI导航自信AI导航
搜索标签

当前位置:首页 > 行业资料 > 其他

移动网页_全站_页脚广告1

关于我们      便捷服务       自信AI       AI导航        获赠5币

©2010-2024 宁波自信网络信息技术有限公司  版权所有

客服电话:4008-655-100  投诉/维权电话:4009-655-100

gongan.png浙公网安备33021202000488号   

icp.png浙ICP备2021020529号-1  |  浙B2-20240490  

关注我们 :gzh.png    weibo.png    LOFTER.png 

客服