1、GE Power&WaterWater&Process TechnologiesAnalytical InstrumentsGE Power&WaterWater&Process TechnologiesAnalytical InstrumentsGE PowerWater&Process TechnologiesAnalytical InstrumentsSievers InnovOx Laboratory TOC Analyzersfor organics monitoringOverviewGE Analytical Instruments Sievers*InnovOx Total O
2、rganic Carbon(TOC)Analyzers are designed to measure organic carbon in a broad range of water samples from industrial process water to wastewater influent and effluent.All InnovOx Analyzers include robust sample handling capability,industry-leading supercritical water oxidation(SCWO)technology,and si
3、mple maintenance and operation,all designed to enhance performance and reliability in challenging applications.For well over 20 years,customers around the globe have trusted the Sievers name for industrial water organic monitoring.Organics Monitoring ApplicationsBiological wastewater plant optimizat
4、ion.Monitoring organics before and after treatment allows operators to optimize the F/M(food to microorganism)ratio,thereby enhancing organic removal,reducing sludge and chemical usage,and avoiding system upsets.Wastewater effluent monitoring and COD/BOD correlation.As an automated laboratory surrog
5、ate for time consuming and manual COD and BOD tests,the InnovOx TOC instrument help monitor the performance of a wastewater system.Optimizing the process provides confidence that regulatory testing results will be below permit limits.High-salt seawater and brine monitoring.GEs patented SCWO oxidatio
6、n technology was designed to measure organics reliably in high-salt or brine samples commonly found in the refining industry.Unlike combustion technologies that can easily be plugged or damaged by salt,the InnovOx oxidation reactor is self-cleaning and is not affected by salts coming out of solution
7、Industrial/commercial laboratory sampling.The InnovOx instrument with optional high-capacity Autosampler is ideal for reliably analyzing water samples from a broad range of sources including water reuse,cooling towers,boilers,research applications,or environmental testing.InnovOx Laboratory TOC Ana
8、lyzer Autosampler Protocol PC Interface BOD or COD values are calculated based on the TOC measurement.Supercritical Water Oxidation(SCWO)Supercritical water oxidation(SCWO)was originally developed to treat large volumes of aqueous waste streams,sludges,and contaminated soils.SCWO destroys organic wa
9、stes using an oxidant in water and temperatures and pressures above the critical point of water:375C(770F)and 22.1 mPa(3,200 psi).These conditions enable rapid and complete oxidation of organic carbon to CO2.Today,SCWO research and development is focused on treating a variety of toxic and hazardous
10、organic wastes.GE Analytical Instruments is the first company to use this technique in a commercial laboratory TOC instrument.Supercritical Water Oxidation(SCWO)Organic+S 2 O 8 -2 +heat+pressure CO2+H2O+SO4-2 Supercritical fluidcritical pointpressureliquidgassolid Supercritical fluid not a gas or a
11、liquidCritical pointTemperature oCPressure(psi)LiquidGasSolid 3200 15 375CNormalBoiling PointGE PowerWater&Process TechnologiesAnalytical InstrumentsProduct CapabilitiesWide dynamic range up to 50,000 ppm TOCAutosampler with stirring and rinse station options for high volume laboratory applicationsP
12、atented supercritical water oxidation(SCWO)for superior TOC recovery and high reliabilityReliable NDIR detection technology with no moving partsStraightforward operation and maintenanceVersatile measurement modes include TOC(TC-IC)or NPOCHandles tough TOC samples,like cellulose and brine,without inc
13、reased maintenance or downtimeComplianceUS EPA Method 415.1 Organic carbon in drinking,surface,seawater,and waste waterUS EPA Method 415.3 Organic carbon in surface and drinking waterUS EPA Method 9060A Organic carbon in ground,surface,saline,and waste waterCEN Method DIN EN 1484 Organic carbon in d
14、rinking,ground,surface,sea,and waste waterISO 8245 Organic carbon in drinking,ground,surface,sea,and waste waterPattern Approval Certificate for Measuring Instruments of the Peoples Republic of China,issued by General Administration of Quality Supervision,Inspection and Quarantine of the Peoples Rep
15、ublic of ChinaOptionsGE Autosampler For high-volume laboratory applications,the InnovOx Analyzer can be used with the GE Autosampler.It handles up to 120 35-mL sample tubes or up to 63 40-mL or 60-mL vials in one protocol.An optional stirring station creates a homogeneous mixture for analysis during
16、 the sampling process and an optional wash station reduces sample-to-sample carryover.Air FilterThe Air Filter is used to purify ambient air and pressurize it sufficiently for use as a sparge and carrier gas.It easily attaches to the back of the analyzer to avoid the use of external carrier gas cyli
17、nders.InnovOx robust SCWO oxidationIndustries Served 2016,General Electric Company.All rights reserved.300 00306 Rev.M.MC15-017 GEA of General Electric Company;may be registered in one or more countries.1 Stated analytical performance is achievable under controlled laboratory conditions that minimiz
18、e operator and standards errors.2 NPOC mode,bottled N2 as carrier gas.This performance is achieved under controlled conditions where variables that influence low-level performance have been minimized or eliminated.This performance may not be achievable on instruments delivered before August 2016.3 U
19、p to 63 sample tubes for 40mL and 60mL vials.Up to 120 sample tubes for 35mL vials.InnovOx Lab TOC Analyzer Operating Specifications1 Analysis Modes NPOC,TOC(TC-IC),TC,IC Range 0.250 to 50,000 ppm TOC Precision RSD 3%of reading at 5 ppm NPOC or TOC Accuracy Greater of 3%or 0.25 ppm,1 to 100 ppm,NPOC
20、 or TOC Linearity R2 0.995,measured as NPOC LOD 0.10 ppm NPOC mode2 TOC Calibration Stability Up to 6 months Analysis Time 2.6 to 8.3 minutes Particle Diameters in Sample 800 m diameter Sample Temperature 10-60 C (41-140 F)Ambient Temperature Range 10 to 40 C (50-104 F)Sampling Ambient pressure,0.08
21、 to 2.82 mL sample volume per analysis Analyzer Specifications Sample Capacity Up to 120 with optional Autosampler3(or single standard vial port)Outputs Ethernet Port(1);USB Ports(3)Display Color LCD w/touch-screen Power 100 240 10%VAC,50 60 Hz,400 VA Installation/Overvoltage Category II Dimensions
22、H:52.05 cm(20.5 in.);W:32.26 cm(12.7 in.);D:58.42 cm(23.0 in)Dimensions(Autosampler)H:52.2 cm(20.5 in.);W:31.1 cm(12.25 in.);D:53.3 cm(21.0 in.)Weight 22.41 kg(49.4 lb);14.1 kg(31.1 lb)for optional Autosampler Safety Certifications CE,ETL listed.Conforms to UL Std.61010-1.Certified to CSA C22.2 NO.61010-1 Environment Maximum Relative Humidity Up to 95%,non-condensing Normal Operating Indoor use only.Keep instrument and reagents out of direct sunlight.Maximum Altitude 3,000 m(9,800 ft)Pollution Degree 2 System SpecificationsGE PowerWater&Process TechnologiesAnalytical Instruments